Laboratory animals, infected with agents assigned to Biosafety Level 4, are housed a Class III biological safety cabinet or in a partial containment caging system (such as open cages placed in ventilated enclosures, solid wall and bottom cages covered with filter bonnets, or other equivalent primary containment systems), in specially designed areas in which all personnel are required to wear one-piece positive pressure suits ventilated with a life support system.

Animal work with viral agents that require Biosafety Level 4 secondary containment, and for which highly effective vaccines are available and used, may be conducted with partial containment cages and without the one-piece positive pressure personnel suit if: the facility has been decontaminated, no concurrent experiments are being done in the facility which require Biosafety Level 4 primary and secondary containment, and all other standard and special practices are followed.